7. Daily Use of the Instrument

4. Another option for removing the blade is to use

the brush with magnet (12).

To do so, fold the clamping lever (10) down-

wards counterclockwise (

Fig. 32

). Fold the

anti-roll guide system (4) to the left.

Guide the brush with magnet to the blade and

lift it upwards and out.

Magnet

12

Fig. 32

10

4

Inserting low-profile blades (continued)

Fig. 30

Two magnets are attached to the rear side of the

guide. These point away from the operator after

the guide has been inserted (towards the rear

pressure plate).
Then insert the blade as described (for high-

profile blade).

Removing the blades

10

9

4

11

1. Fold the anti-roll guide system (4) to the

left – while doing so, hold the lever (11) (not

the adjusting screw of the anti-roll guide) so

that the height of the anti-roll guide remains

unchanged.

2. Open the clamping lever (10) by turning it

counterclockwise (

Fig. 31

).

3. Carefully lift out the blade (9). Continue with

instruction 5.

Fig. 31