Operation – Leica Biosystems VT1000 A User Manual

Page 32

Advertising
background image

32

Operating Manual V 1.1 – 01/2010

Fig. 47

Adjustable sectioning window

(continued)

• First, adjust the front limit switch that is on

the front side of the instrument.

• To do so, release the knob (57.

1

) with the

blade fully retracted and move it into the de-

sired position.

• As soon as the switch is in the desired posi-

tion, tighten the knob again so that it cannot

be disengaged during sectioning.

• The rear switch can be moved only while the

instrument is in operation.

• To adjust the rear limit switch, turn the knob

(57.

2

) counterclockwise to disengage the

switch.

• Then, start the sectioning operation using

the switch for the feed movement of the Vi-

bratome.

• During the feed movement of the blade,

push the switch forward until it reaches the

desired position.

• Finally, tighten the knob to fix the switch in

place.

• You can readjust the sectioning window

whenever new specimen sizes are used.

To adjust the front and rear edge of the section-

ing window, you have to move the internal limit

switch using two rotary knobs (57), which are

on the side of the top plastic cover of the Vi-

bratome. The two rotary knobs can be adjusted

independently of one another.

57.

2

To make this adjustment, the instrument should

be in manual mode.

57.

1

5.

Operation