Leica SP 2600 – Ultramilling attachment

7.2.2 Replacing the light bulb in the calibration device

• Remove base cover (2) of calibration device (1) (unscrew 4 slotted

screws (3)).

• Loosen light bulb holder (5) (metal bridge with lamp socket) (unscrew 2

slotted screws (4)).

• Take replacement light bulb (8) out of sponge rubber cube (7).
• Remove broken light bulb (6) from lamp socket and replace with new

light bulb (8).

• Place another replacement bulb (8) into the sponge rubber cube (7).
• Put light bulb holder (5) back in place.
• Place ring with cable (9) onto bore (10) in the light bulb holder.
• Fasten light bulb holder with two screws (4).
• Put sponge rubber cube (7) with new replacement light bulb (8) back into

the calibration device.

• Put base cover (2) back in place and fasten with 4 screws (3).

7. Cleaning, disinfection and maintenance