7. Troubleshooting

Problem

Possible cause

Corrective action

5. Illumination

does not work.

- Button on the head of the lamp

has not been pressed.

- Bulb defective.

- Fuse defective.

- Press the On/Off button on the

head of the lamp.

- Ask technical service to re-

place the lamp.

- Change fuse.

(See

Chapter 7.3

)

6. Individual heating areas

(Despite correct temperature indi-

cation) heating does not work.

- Fuse for this heated area has

responded.

- Heater defective.

- Change fuse.

(see

Chapter 7.3

)

- Contact customer service.

7. Error message in display


The

message appears

in the display.

Program and time are lost.

- The battery is empty

and also

- Power failure or

- Device switched off with stand-

by switch or

- Power plug disconnected.

- Acknowledge fault by pressing

any key.

Work can resume after repro-

gramming.

Important!

Do not disconnect the instru-

ment from the power supply

anymore.

Ask technical service to replace

battery.