Safety – Leica Biosystems RM2265 User Manual

Page 9

Advertising
background image

7

Leica RM2265

Personal safety precautions

2. Safety

Warnings – Safety notes on the instrument itself

Warnings – Transport and installation

• Safety notes on the instrument itself, which are marked with a warning triangle, indicate that

the correct operating instructions (as defined in these Instructions for Use) must be followed

when operating or replacing the item marked.

• Failure to adhere to these instructions may result in an accident, personal injury, damage to

the instrument or accessory equipment.

• Once unpacked, the instrument may be transported only in an upright position.
• Never lift the instrument by the handwheels or the cassette clamp. Always remove the sec-

tion waste tray before transporting the instrument.

• Caution! The voltage selector has been preset at the factory.
• Before connecting the instrument to the line voltage, please check that this setting complies

with the local power requirements of your laboratory.

• The mains cable inlet is closed with an adhesive tape that indicates the actual voltage set-

ting of the instrument upon arrival. Severe damage can be caused to the instrument if the

voltage selector is set to an incorrect voltage!

• When adjusting the voltage selector setting, the instrument must be disconnected from the

mains.

• The instrument must be connected to mains only with one of the supplied power cords and

only to grounded sockets. Do not interfere with the grounding function by using an extension

cord without a ground wire.

• The instrument is ready for operation only if a dummy plug or foot switch (optional) is con-

nected. If you hear a beeping sound (continuous) after switching on the instrument, check

that the control panels are connected correctly.

• The instrument may not be operated in hazardous locations.
• Extreme temperature fluctuations between storage facility and setup site as well as high

humidity may cause condensation to form. If this is the case, wait at least two hours before

switching on the instrument. Failure to comply with this may cause damage to the instrument.

• Protective devices on both instrument and accessories must neither be removed nor modified.

• When working with microtomes, personal safety precautions must always be taken. It is

mandatory to wear work safety shoes, safety gloves, a mask and safety goggles.