45

Leica ST4040 – Linear Stainer

8.

Cleaning and Maintenance

8.1

Cleaning

Clean the reagent and tap water stations as well as the trough which
holds the reagent and tape water stations regularly.

For that purpose, remove the reagent and tap water stations from the
trough.

The reagent and tap water stations can be washed in a laboratory
dishwasher.

Wash the tap water and reagent stations in the dishwasher at a tem-
perature of max. +65 °C . - Use a standard detergent for laboratory
dishwashers.
At any rate avoid washing the stations at higher temperatures (e. g.
in industrial diswashers which run at a temperature of +85 °C), as the
stations may become deformed!

To clean the painted instrument surfaces and the lids, use a mild
household detergent (see safety instructions on previous page for non-
appropriate ingredients).

From time to time check the drain hose for accumulated dirt, especially
algae - clean if necessary.