4 installing the oven (optional), Installing the oven (optional), Setting up the instrument – Leica Biosystems ST5020 User Manual

Page 21

Advertising
background image

21

Leica ST5020

4.4 Installing the oven (optional)

The following parts must be in-
cluded:

15 - Oven module

16 - Paraffin collecting tray

17 - Cover

18 - Screws

19 - Washers

20 - Screwdriver

21 - Allen key, size 2.5

22 - Allen key, size 4.0

Take the oven kit out of the box and verify
whether it is complete.

15

16

17

18

19

22

21

20

Fig. 10

Insert paraffin collecting tray
into corresponding opening in
oven module (possible only in
one direction, as shown here).

Check the voltage selector setting and ad-
just if necessary.

Important!
The voltage selector switch must be
set to the same mains voltage as the
instrument.

To adjust the voltage selector setting, insert
screw driver (20) into gap (23) and turn care-
fully until the white point above the correct
voltage indication is located underneath the
white triangle.

115 means: Instrument voltage 100 -120 V.

230 means: Instrument voltage 230 - 240 V.

Fig. 11

Rear panel - oven

Voltage setting

100 - 120 V

Voltage setting

230 - 240 V

24

25

23

4.

Setting up the instrument