Leica MC120 HD / MC170 HD User Manual

Application 32

General Notes

 

The white balance ensures that the
specimen is shown in neutral colors. We

recommend carrying out a new white balance
adjustment whenever you change the illumina-
tion type or color temperature. This is particu-
larly necessary if you are working with halogen
light that can be switched from yellow (low
intensity) to blue (high intensity).

 

Configure the white balance via the user
menu using the remote control. Depend-

ing on the configuration, automatic white
balance can also be carried out using the foot-
switch.

 

The white balance of the camera is
factory set to automatically provide opti-

mum results when used in conjunction with a
Leica LED illuminator.

Manual white balance

O

Manual white balance is described on
page 38.

Automatic white balance
1. Position a gray chart or a gray, neutral

object under the microscope so that the
entire image field is filled in.

2. Adjust the illumination as desired.

3. Press the button

on the remote control.