5.2 Procedure

The primary tilt of the instrument can be in either the ML or AP direction so there are two

procedures for the vertical alignment. Select the correct instruction set for your instrument set up.

Set the Rotation and Tilt

The instrument’s rotation and tilt must be moved so that the scales read zero.

1. Unlock the tilt and rotation clamps, and move the instrument so the scales read zero, as shown

in Figure 35.

2. Relock the clamps when finished.

Figure 35. Both the tilt and rotation are at zero

Installing the Alignment Kit

The alignment kit needs some assembly before being installed.
3. Double-click the ‘Angle Two’ icon on the desktop to start the program.
4. Remove the ear bars and probe shaft from the instrument.

Zero tilt

Zero rotation

Tilt clamp

Rotation clamp