2 procedure, Procedure – Leica Biosystems ANGLE TWO User Manual
Page 34
Alignment
Leica Angle Two
™
User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010
34
5.2 Procedure
The primary tilt of the instrument can be in either the ML or AP direction so there are two
procedures for the vertical alignment. Select the correct instruction set for your instrument set up.
Set the Rotation and Tilt
The instrument’s rotation and tilt must be moved so that the scales read zero.
1. Unlock the tilt and rotation clamps, and move the instrument so the scales read zero, as shown
2. Relock the clamps when finished.
Figure 35. Both the tilt and rotation are at zero
Installing the Alignment Kit
The alignment kit needs some assembly before being installed.
3. Double-click the ‘Angle Two’ icon on the desktop to start the program.
4. Remove the ear bars and probe shaft from the instrument.
Zero tilt
Zero rotation
Tilt clamp
Rotation clamp