Ich 500, Setup, Operation and control co – Memmert ICH C (with CO2 control) Climate chambers User Manual

Page 29: Timer end

Advertising
background image

D30393 | Date 10/2014

29

Operation and control

CO

2

Adjustment range: 0 to 10 % or 0 to 20 % in steps of 0.1% (de-

pends on appliance size)

CO2

1.

0

%

15.

0

%

Set

5.4.5 Operation with digital backwards counter with target time setting,

adjustable from 1 minute to 99 days (Timer)

In timer operation, you can adjust the time the appliance runs at the set value. The appliance

has to be in manual operating mode for this.

1. Press the activation key to the left of

the timer display. The timer display is

activated.

30.0

%rh

50.0

%CO

2

Programm 12

Fr 20.10.2010

20:31

min

000°C

max

000°C

.5°C

100

ICH 500

MENU

30.0

%rh

50.0

%CO

2

Programm 12

Fr 20.10.2010

20:31

min

000°C

max

000°C

MENU

TIMER

-

Ende 9:00 23.11.

--

h

-

m

2. Turn the turn control until the desired

duration is displayed – in this example 4

hours 30 minutes. The approximate end

time is shown beneath, in a smaller font.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

TIMER

End

0

13:30 23.11.

04

m

h

3

Up to a duration of 23 hours 59 minutes, the time is displayed in hh:mm (hours:minutes)

format. For 24 hours and more, the format dd:hh (days:hours) is used. The maximum

duration adjustable is 99 days 00 hours.

3. Press the confirmation key to confirm.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

The display now shows the remaining time

in a large font and the approximate end time

in a smaller font beneath. The status display

shows „Timer active“.

TIMER

30

m

04

h

Ende 13:30 23.11.

Timer active

12.Sept.2012 13:44

4. Now, as described under 5.4.4, set the individual values for temperature, air flap position

etc. which you want the appliance to operate at. The set values can be changed at any

time during the timer. The changes are effective immediately.
In

Setup

, you can choose if the timer should run setpoint-dependent or not. This deter-

mines whether the timer should not start until a tolerance band around the set tempera-

ture is reached or if it should start right after activation (see page 49). If the timer runs

setpoint-dependent, this is indicated by the

symbol in the timer display.

Advertising