23 Temperature monitor and protection devices

The monitor temperature is measured with the Pt100 temperature sensors in the heating trays.
The monitoring system serves as protection for the oven load and protects the oven and its surroundings.

The oven is provided with duplicate overtemperature protection (mechanical / electronic) according to DIN 12 880.

MIN

MAX

AUTO

°C

23.1 Mechanical temperature monitor: temperature limiter (TB)

All ovens of the VO series are equipped with a mechanical temperature limiter (TB) Protection Class1 to
DIN 12 880.
If the electronic monitor system should fail during operation and the fixed factory-set maximum temperature
is exceeded by approx. 20°C the temperature limiter switches off the heating permanently as a final
protective measure. The alarm symbol lights up as warning

Fault rectification after the TB cut-out has been activated:
1.

Switch off the oven and allow it to cool down

2.

Rectify the fault (e.g. replace temperature probe) and where appropriate contact customer service

3.

The oven is again ready for operation only after it has cooled down and after the fault has been

rectified

Overtemperature

protection

(TWW, TWB)

Undertemperature protection

Audible

alarm signal

Automatic

temperature monitor

(ASF)

visual alarm symbol
alight: TB alarm
flashing:

TWW alarm

TWB-alarm

ASF-alarm