Going outside tolerance margin = asf alarm – Memmert IFP 400 - 800 Incubators (Generation 2003) User Manual

Page 30

Advertising
background image

PERFECT

page 30

Going outside tolerance margin = ASF alarm

Going outside the selected tolerance margin of the setpoint (in the example 150°C +/-10°C), for example
through opening the oven door during operation, triggers the alarm.
Triggering the ASF alarm is indicated by flashing

AUTO

and

-symbol.

If the sounder is switched on in SETUP, the ASF alarm is additionally signalled by an interrupted tone. By
pressing the SET key the sounder can be switched off temporarily until the next occurrence of an alarm
event.

160°C

140°C

ASF alarm

ASF activated

ASF activated

160

140

ASF-alarm

Again within tolerance margin = ASF alarm switched off

The automatic temperature monitor alarm is switched off automatically as soon as the selected tolerance
margin of the setpoint (in the example 150°C +/-100°C) is reached again.

160°C

140°C

ASF alarm

ASF activated

ASF activated

160°C

140°C

ASF-alarm

ASF activated

ASF set to

+/-10°C in

SETUP

ASF set to

+/-10°C in

SETUP