Steps to Run a Test
MTS Series 793 Controller Overview
Perform a Practice Test
93
7. Set up the test monitor.
Set up the Meters and Scope as desired to monitor test signals.
8. Set up a test program.
A. From the Application control bar, select Basic TestWare.
B. In the Command tab, set the following:
Note
Ensure that the Target Setpoint and Amplitude values are appropriate for
your particular specimen.
C
ONTROL
S
ETTING
Type
Cyclic
Channel
Channel 1
Control Mode
Force
Target Setpoint
Guideline: 25% of full scale
Amplitude
Guideline: 10% of full scale
Frequency
1 Hz
Wave Shape
Sine Tapered
Compensator
None
Start Action
Disabled
Done Action
Disabled