Fundamental Video Extensometer

24

Starting the Video Extensometer

9. The buttons:

Press the Save INI… button to record the test configuration inside a file
in the selected directory.

Press the OK button to accept the parameters and proceed to the Test
window. Or in the case of bi-camera configuration, proceed to the
Settings window for the second camera.

Press the Cancel to quit the application without saving changes.

Press the Files DCF… button to select the configuration file of the
camera (Full resolution, Partial Scan ½, 1/3, bi-camera).

Available DCF files are:

METAL_FULL_SAP.dcf or METAL_FULL_BC_SAP.dcf

Allows you to carry out tests with the full resolution of the
camera; The number of images which can be processed is 16
images/second with a resolution of 1380 * 1035.

METAL_PAR2_SAP.dcf or METAL_PAR2_BC_SAP.dcf

Allows you to carry out tests by reducing the width of the image
by ¼ on each sides; The number of images which can be
processed is 30 images/second with a resolution of 1380 * 513.

METAL_PAR3_SAP.dcf or METAL_PAR3_BC_SAP.dcf

Allows you to carry out tests by reducing the width of the image
by ⅓ on each sides; The number of images which can be
processed is 43 images/second with a resolution of 1380 * 341.