Steps to run a test, Steps to run a test 40 – MTS 85X Damper User Manual

Page 40

Advertising
background image

Damper System Operation

40

Perform a Practice Test

793 Controller Operation Overview

Tune the system

To perform the Series 793 Software installation, see the Installation Instructions
readme file on the Series 793 System Software CD, and follow the installation
wizard.

To perform the remaining tasks, see the MTS Series 793 Tuning and Calibration
manual.

You may also find it necessary to see the following manuals:

MTS Series 793 Utility Software manual—To setup or edit the hardware
interface file to reflect any hardware changes to your system.

MTS Series 793 Control Software manual—To find detailed information
about using Series 793 Software controls.

Other MTS manuals—You can receive additional MTS manuals that
provide instructions for maintaining and operating your test system.

Steps to Run a Test

Improper controller setup can result in unexpected actuator movement.

Unexpected actuator movement can injure personnel and damage test
equipment.

Ensure that your controller is set up properly before attempting to perform a test.

1. Power up the controller chassis and PC.

2. Start the Station Manager application and select a Project.

A.

From your desktop, start the Station Manager application:

Start > Programs > MTS 793 Software name (for example, FlexTest
GT
) > Station Manager

B.

As the Station Manager application loads, you will be prompted to
select a Project. Select the existing Project that was created for your
system.

3. Apply hydraulic pressure.

When you power up your controller, the hydraulic interlock will be set.
Perform the following on the Station Controls panel:

A.

Clear system interlocks.

Click Reset. If the interlock indicator lights again, determine the cause
of the interlock and correct it before proceeding.

B.

If HPU controls are present, click the

(Low) and then

(High) buttons.

WARNING

Advertising