Pump stalls and is unable to turn: This may be because the pumping rate is set too high. Reduce rate.
Peristaltic tubing being used may be too rigid for this pump. Replace tubing.
Liquid may be too viscous or there may be too much pressure on the system.

There may be too much tubing in the pump head, reduce distance between tie-raps and recalibrate pump,
refer to section 4.2.1.

Pump doesn’t stop after dispensing a set volume: The pump was previously setup with a multiple
Phase Pumping Program. To simply dispense a fixed volume at a fixed pumping rate, the second Program
Phase must be the ‘Stop’ function. See section 11.1, How to Enter Pumping Programs, for instructions on
changing the Pumping Program.

Pump stops pumping after a period of time: A dispense volume has been set. Verify that the dispense
volume is set to 0.

14.5

Specifications

14.5.1

Mechanical & Electrical

Number of pumps:

1

Motor type:

Step motor

Motor steps per revolution:

200

Microstepping:

1/8 to 1/1 depending on motor speed

DC connector:

2.1mm, center positive

Voltage at DC connector:

24V DC at full load

Amperage:

900mA at full load

Power supply type:

Unregulated linear external wall adapter, country and power
source specific

Power supply output rating:

24V DC @ 1A

Dimensions:

7 ¾” x 5 1/4” x 5 3/4” High
(19.685 cm x 13.335 cm x 14.605 cm)

Weight:

4.51 lbs. (2.05 kg)

14.5.2

Operational

Maximum speed:

372 rpm

Minimum speed:

0.0168 rpm

Maximum pumping rate:

775.2 mL/min with 3/16 ID tubing

Minimum pumping rate:

0.035 mL/min with 3/16 ID tubing

Number of Program Phases:

40

RS-232 pump network:

100 pumps maximum

RS-232 selectable baud rates:

300, 1200, 2400, 9600, 19200

Tubing inside diameter range:

Below 1” diameter

14.6

Custom Applications

For specialized and OEM applications, contact your dealer
or NEW ERA PUMP SYSTEMS INC.. Custom
modifications can be made to the mechanics or the
firmware.








Advertising