Model NE-1600/NE-1800 Multi-Phaser™

Publication #1200-02

38

8/1/14

PURGE PUMP
PUR

Starts purge. Pump infuses or withdraws at the top speed, depending on the pumping direction.

STP command or ‘Start/Stop’ key will stop the pump.

STOP PUMPING PROGRAM
STP

If the Pumping Program is operating, the pump will be stopped and the Pumping Program will be
paused.

If the Pumping Program is paused, the stop command will cancel the pause and reset the Pumping
Program.

VOLUME DISPENSED

DIS

Queries volume dispensed only. Set not applicable.
Response:

I <float> W <float> <volume units>

Where: “I <float>“ refers to the infusion volume dispensed, and “W <float>” refers to the withdrawn
volume.

CLEAR VOLUME DISPENSED

CLD { INF | WDR }

Sets the Infused or withdrawn volume dispensed to 0. Command is only valid while the Pumping
Program is not operating. Query is not applicable.

INF

= Infusion volume

WDR

= Withdrawn volume

Query is not applicable.

9.4.3

Configuration and Setup Commands

New settings for any of the following commands will be stored in the non-volatile memory.

LOW MOTOR NOISE
LN [ <on-off> ]

Set/query low motor noise mode. Increases motor stepping resolution.

ALARM SETUP
AL [ <on-off> ]

Set/query alarm setup mode. Set alarm enables or disables alarm buzzer mode.

POWER FAIL SETUP
PF [ <on-off> ]

Set/query Power Failure mode. Set Power Failure enables or disables Power Failure mode.

TTL I/O OPERATIONAL DEFAULT TRIGGER SETUP
TRG [ <trigger setup> ]

Set/Query TTL I/O Operational Trigger (Pin 2) default input configuration.

TTL I/O Operational Trigger is set to <trigger setup>.

<trigger setup> =>

FT

= Foot switch trigger (falling edge start/stop)

FH

= Foot switch hold (falling edge start, rising edge start)

F2

= Foot switch reverse (rising edge start/stop)

LE

= Level trigger (rising edge start, falling edge stop)

ST

= Start only trigger (falling edge start)

T2

= Start only trigger reversed (rising edge start)

SP

= Stop only trigger (falling edge stop)

P2

= Stop only trigger reversed (rising edge stop)

RL

= Start on low level

RH

= Start on high level

SL

= Stop on low level

SH

= Stop on high level