New Era NE-4000 Multi-Phaser Programmable Double Syringe User Manual

Page 44

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-4000 Multi-Phaser™

Publication #1200-01

39

09/05/14


BUZZER

BUZ [ 0 | { 1 [ < n > ] } ]

Sets / queries buzzer

Set:

0 = Turn buzzer off;

1 = Turn buzzer on

if <n> specified

If <n> = 0, buzzer beeps continuously,

otherwise buzzer beeps <n> times

if <n> not specified, buzzer sounds continuously

Query response: { 0 | 1 }

0 = Buzzer off
1 = Buzzer is on continuously or beeping.

9.4.5

System Commands

SET PUMP NETWORK ADDRESS AND BAUD RATE

* ADR [ <address> [ B { 19200 | 9600 | 2400 | 1200 | 300 } ] ]
[DUAL | RECP | ALTR]

Set/query pump network address

<address> => <n> [<n>]
<address> Valid range: 0 to 99

B { 19200 | 9600 | 2400 | 1200 | 300 } will change the baud rate as indicated. NOTE: The command

response and all further communications will be at the specified baud rate.

Special communications modes

DUAL Sets the pump to Dual Pumping mode with a secondary pump.
RECP Sets the pump to Reciprocating Pumping mode with a secondary pump.
ALTR Sets Alternating pumping mode.

All special communications modes use 19,200 baud rate.

This is a special system command that will be accepted by the pump regardless of its current address or mode.
Once set, the pump will only respond to commands with the set address and at the specified baud rate.

Note: Once a special communications mode is set, the pump will only respond to commands that are preceded by
the ‘*’ character. To exit special communication mode, reset the address: * ADR 0

Example: *ADR

Query current address setting

*ADR 3

Set pump network address to 3. The pump will now only respond to
commands with address 3

*ADR 5 B 1200

Set the pump network address to 5 and the baud rate to 1200. The command
response will be at 1200 baud

ENABLE SAFE COMMUNICATIONS MODE

SAF [ <time out> ]

Set/query Safe communications mode setting.
<time out> => <n> [ <n> [ <n> ] ]
<time out> Valid range: 0 to 255.

If <time out> = 0 then Basic communication mode is set, disabling Safe mode,

If <time out> > 0 then Safe communications mode is enabled. After the reception of this command, valid
communications must be received every <time out> seconds.