New Era FW-1-X FIRMWARE UPGRADE NE-1000X FAMILY OF SYRINGE PUMPS User Manual

Page 5

Advertising
background image

NE-1000X

5

www.SyringePump.com








Phase

Function

Rate

Time

Direction

8

LINR

0 ml/hr

00:10

Infuse

Phase

Function

Rate

Time

Direction

9

LINR

1000 ml/hr

00:00

Infuse

Phase

Function

Rate

Time

Direction

10

LINR

1000 ml/hr

00:10

Infuse

Phase

Function

Rate

Time

Direction

11

LINR

0 ml/hr

00:00

Infuse

Phase

Function

12

RET



The above program would be entered from RS-232 with the following commands:

PHN 1
FUN RAT
RAT 100 MH
VOL 1
DIR INF

PHN 2
FUN INC
RAT 200
TIM 00:10
DIR INF

PHN 3
FUN JMP 08

PHN 4
FUN RAT
RAT 500 MH
VOL 5.0
DIR INF

PHN 5
FUN DEC
RAT 100.0
TIM 00:10
DIR INF

PHN 6
FUN JMP 08

PHN 7
FUN JMP 01

PHN 8
FUN LIN
RAT 0.0
TIM 00:10
DIR INF

PHN 9
FUN LIN
RAT 1000 MH
TIM 00:00
DIR INF

PHN 10
FUN LIN
RAT 1000 MH
TIM 00:10
DIR INF

PHN 11
FUN LIN
RAT 0.0 MH
TIM 00:00
DIR INF

PHN 12
FUN RET













Return to Phase following calling
Program function

Linearly ramp the infusion rate
from 1000 ml/hr to 0 ml/hr over 10
minutes

Linearly ramp the infusion rate
from 0 ml/hr to 1000 ml/hr over 10
minutes

Linear Ramping Sub-Routine

Advertising