Model 525 Operating Instructions

82 E. Main Street Suite 3.14 · Webster, NY 14580

Tel: 585-872-9350 · Fax: 585-872-2638

[email protected] · http://www.piecal.com/

Practical Instrument Electronics, Inc. Copyright ã 2011. All rights reserved.

525-9002 7/13/11

3-8

Source Mode

Slide the SIMULATE/OFF/READ switch to SIMULATE for direct RTD or T/C output. The Model 525 outputs resistance corresponding to

temperature for the selected RTD or outputs mV corresponding to temperature for the selected T/C type.
Turn the EZ-Dial™ Knob to change temperature, push and turn for faster dialing.
Slide the EZ-Check™ Switch to HI or LO to recall stored settings. While in the HI or LO position, dial a new setting and press the

EZ-Dial™ Knob to store. The DIAL position always holds the last setting dialed there. Using the STORED Values gives you very

accurate values for repeatability. Adjusting the EZ-Dial™ Knob you can check for controller actions, trip points or hysteresis.
Double-click the EZ-Dial™ Knob to return to the configuration menu.

Auto Stepping in Source Mode

Select the step size by pressing the EZ-Dial™ Knob to find the desired step size or %. Select the step time by pressing the EZ-Dial™

Knob to find the desired step time indicated in seconds.

RTD Connection Diagrams, (wire lead kit supplied)

SOURCE ONLY: (See opposite figure)
To connect to a 2 wire transmitter connect one red lead to

post 1 (RTD1) and connect one black lead to post 2 (RTD2).
To connect to a 3 wire transmitter connect one red lead to

post 1 (RTD1) and connect two black leads to post 2 (RTD2).
To connect to a 4 wire transmitter connect two red leads to

post 1 (RTD1) and connect two black leads to post 2 (RTD2).

NOTE:
3 WIRE & 4 WIRE TERMINALS ARE FOR READ MODE