Practical instrument electronics, Calibrating rtd instruments, Reading rtd sensors – Practical Instrument Electronics 311 User Manual

Page 4: While the ez-check™ switch

Advertising
background image

Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 4

SOURCE

Choose this function to provide a simulated RTD signal into controllers, temperature transmitters,
indicators or any input devices that measure thermocouple sensors.

1) Disconnect the RTD sensor from the device to

be calibrated.

2) Select “

SOURCE” with slide switch

w

.

3) Connect the PIECAL 311 to the device using 2,

3 or 4 wires matching the connections of the
sensor that was just removed.

The output is adjusted in 0.1° (or 0.01/0.1 ohm)
increments by turning the knob

e

while the

EZ-CHECK™ switch

q

is in the “HI”, “LO” or “SET”

position. Press and turn the knob for faster dialing
with 10° (or 1.00/10.0 ohm) increments.

The OVERLOAD indicator will light if excessive
voltage or current is detected by the calibrator.

Calibrating RTD Instruments

3 2 2 - 1

THERMOCOUPLE

CALIBRATOR

J, T, E, K, mV

HI

Pt100 α=385

AUTOMATED

UNIVERSAL RTD

CALIBRATOR

READ

SOURCE

HI

SET

LO

MAX

READ

MIN

OVERLOAD

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

OFF

°C

138.50Ω

3W 2W 1W 4W

  

100.0

READ

Choose this function to measure temperatures with an RTD probe or sensor.

1) Disconnect the RTD sensor from any other device.
2) Select “

READ” with slide switch

w

.

3) Place the EZ-CHECK switch into the READ position.
4) Connect the PIECAL 311 to the device using 2, 3 or 4 wires.

The PIECAL 311 measures the temperature signal and constantly updates the display with the
current temperature reading. Move the EZ-CHECK switch to MAX to see the highest temperature
reading and to MIN to see the lowest temperature reading. Press and hold the knob

e

to clear the

MAX and MIN readings.

The OVERLOAD indicator will light if excessive voltage or current is detected by the calibrator.

Reading RTD Sensors