Ranges & Accuracies

T/C

Degrees C

Range

°C

Degrees F

Range

°F

T/C

Material

J

-200.0 to -180.0

±0.3°

-346.0 to -292.0

±0.5°

+Iron

-Connstantan

Jacket

-180.0 to -50.0

±0.2°

-292.0 to -58.0

±0.4°

-50.0 to 500.0

±0.1°

-58.0 to 932.0

±0.2°

500.0 to 1200,0

±0.2°

932.0 to 2192.0

±0.4°

K

-230.0 to -100.0

±0.6°

-382.0 to -148.0

±1.1°

+ Chromel®

-Alumel®

Jacket

-100.0 to 1050.0

±0.2°

-148.0 to 1922.0

±0.4°

1050.0 to 1371.1

±0.3°

1922.0 to 2500.0

±0.5°

T

-260.0 to -200.0

±1.0°

-436.0 to -328.0

±1.8°

+Copper

-Constantan

Jacket

-200.0 to -50.0

±0.5°

-328.0 to -58.0

±0.9°

-50.0 to 0.0

±0.2°

-58.0 to 32.0

±0.4°

0.0 to 400.0

±0.1°

32.0 to 752.0

±0.2°

E

-240.0 to -200.0

±0.4°

-400.0 to -328.0

±0.7°

+Chromel

-Constantan

Jacket

-200.0 to -100.0

±0.2°

-328.0 to -148.0

±0.4°

-100.0 to 850.0

±0.1°

-148.0 to 1562.0

±0.2°

850.0 to 1000.0

±0.2°

1562.0 to 1832.0

±0.4°

General
Accuracy

±(0.008% of Reading + 0.006 mV)

Cold Junction Compensation ± 0.16°F (±0.1 °C)
Millivolt Range

-13.000 to 80.000 mV

Operating Temp. Range

-25 to 60 °C (-10 to 140 °F)

Relative Humidity Range

10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing

Size

L=5.63 x W=3.00 x H=1.60 inches

Weight

12.1 ounces (including boot & batteries)

Batteries

Four “AA” Alkaline 1.5V (LR6)

Battery Life

50 Hours; Low Battery indication with 1 hour life left

Optional NiMh

Rechargeable battery kit

120 VAC for North America Only; charger, four

NiMh batteries, AC & DC cords [Part # 020-0103]

Protection against

misconnection

Over-voltage protection to 60 V dc (rated for 30

seconds)

Display

High contrast graphic liquid crystal display. LED

backlighting for use in low lit areas.

Source
Output Impedance

< 0.3 Ohms

Source Current

> 20 mA (drives 80 mV into 10 Ohms)

Noise

≤ 4 microvolts p-p for frequencies of 10 Hz or below

Read
Input Impedance

> 10 Megohms

Open Thermocouple

Threshold
Pulse


10,000 Ohms nominal
< 10 microamp pulse for 300 milliseconds

Normal Mode Rejection

50/60 Hz, 50 dB

Common Mode Rejection

50/60 Hz, 120 dB

Accessories

Included:

Rubber Boot, Four “AA” Alkaline batteries, Certificate of Calibration

Small Carrying Case with PIE Logo Part No. 020-0205

Optional:

Ni-MH 1 Hour Charger with 4 Ni-MH AA Batteries Part No. 020-0103

(100-120 V AC input for North America Only)

T/C Wire Kit 1 for Types J, K, T & E Part No. 020-0202

T/C Wire Kit 2 for Types B, R/S & K Part No. 020-0203

Three feet (1 meter) of T/C extension wire, stripped on one end with

a miniature T/C male connector on the other end.

Warranty

Our equipment is warranted against defective material and

workmanship (excluding batteries) for a period of three years
from the date of shipment. Claims under warranty can be made
by returning the equipment prepaid to our factory. The equipment
will be repaired, replaced or adjusted at our option. The liability
of Practical Instrument Electronics (PIE) is restricted to that
given under our warranty. No responsibility is accepted for
damage, loss or other expense incurred through sale or use of
our equipment. Under no condition shall Practical Instrument
Electronics, Inc. be liable for any special, incidental or
consequential damage.

Additional Information

PIE Calibrators are manufactured in the USA. This product is

calibrated on equipment traceable to NIST and includes a
Certificate of Calibration. Test Data is available for an additional
charge.

Practical Instrument Electronics recommends a calibration

interval of one year. Contact your local representative for
recalibration and repair services.

P

ractical

I

nstrument

E

lectronics

841 Holt Road #1 • Webster, NY 14580

Tel: 585.872.9350 • Fax: 585.872.2638

[email protected] • www.piecal.com

Copyright © 2011 All rights reserved 322-9001 - Rev D 19 July 2011

PIECAL 322 Specifications

(Unless otherwise indicated all specifications are rated from a nominal 23°C, 70% RH for 1 year from calibration)

Available From:

Advertising