Practical instrument electronics, Thermocouple ranges & accuracies – Practical Instrument Electronics 322-1 User Manual

Page 6

Advertising
background image

Page 6

Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

T/C

Type

Degrees C

Range

Accuracy

Degrees F

Range

Accuracy

T/C

Material

ISA/ANSI

Color

J

-200.0 to -180.0

±0.3°

-346.0 to -292.0

±0.5°

+Iron

-Connstantan

Jacket

White

Red

Black

-180.0 to -50.0

±0.2°

-292.0 to -58.0

±0.4°

-50.0 to 500.0

±0.1°

-58.0 to 932.0

±0.2°

500.0 to 1200,0

±0.2°

932.0 to 2192.0

±0.4°

K

-230.0 to -100.0

±0.6°

-382.0 to -148.0

±1.1°

+ Chromel®

-Alumel®

Jacket

Yellow

Red

Yellow

-100.0 to 1050.0

±0.2°

-148.0 to 1922.0

±0.4°

1050.0 to 1371.1

±0.3°

1922.0 to 2500.0

±0.5°

T

-260.0 to -200.0

±1.0°

-436.0 to -328.0

±1.8°

+Copper

-Constantan

Jacket

Blue

Red

Blue

-200.0 to -50.0

±0.5°

-328.0 to -58.0

±0.9°

-50.0 to 0.0

±0.2°

-58.0 to 32.0

±0.4°

0.0 to 400.0

±0.1°

32.0 to 752.0

±0.2°

E

-240.0 to -200.0

±0.4°

-400.0 to -328.0

±0.7°

+Chromel

-Constantan

Jacket

Purple

Red

Purple

-200.0 to -100.0

±0.2°

-328.0 to -148.0

±0.4°

-100.0 to 850.0

±0.1°

-148.0 to 1562.0

±0.2°

850.0 to 1000.0

±0.2°

1562.0 to 1832.0

±0.4°

Thermocouple Ranges & Accuracies

Based on ≤ ± (0.008 % of Reading + 0.006 mV)

Note: Doesn’t include cold junction error of ±0.05°C