RKI Instruments Eagle 1 Quik Start User Manual

Quick reference guide for the eagle

Advertising
background image

Quick Reference Guide For The Eagle

.

Turn on and adjust the Eagle gas monitor in a known fresh air area

1. Turning the Eagle ON

a. Connect the hose and probe to the Eagle inlet fitting.
b. Press and hold the POWER ENTER button for a few seconds

to turn the Eagle on.

c. Allow Eagle warm up. If Lunch Break feature is active, follow

on-screen prompts.

d. The Eagle will display minimum battery voltage of 4.5 V and

current battery condition.

e. If Data Logging option is installed the Eagle will display Date

and Time.

f. Self Diagnosis will be displayed.
g. The Eagle will sound a double beep and display all gas

readings.

2. Sensor Failure

a. If the Eagle experiences a sensor failure during start up, the

LCD will indicate FAIL SENSOR and indicate the sensor that

has failed in brackets. Example: FAIL SENSOR <OXY>. The

Eagle alarm will sound to indicate failure, then indicate SELF

DIAGNOSIS, beep twice, then display gas readings. OXY (or

failed sensor) will be displayed as xxx’s.

b. If a sensor should fail, replace sensor as required and

calibrate instrument before use.

3. Performing a Demand Zero

a. Find a fresh-air environment free of toxic or combustible

gases and of normal oxygen content (20.9%).

b. Press and hold the AIR button. The display will indicate

DEMAND ZERO, HOLD AIR KEY, ADJUSTING ZERO,

READING ZERO DATA, ZERO ADJUSTED. Release AIR

button when RELEASE AIR KEY is displayed. The Eagle will

set to fresh air values, (0% LEL, 20.9 %Vol O2, 0.0 ppm H2S

and 0 ppm CO).

4. Performing a Flow Integrity Test

a. The flow integrity test verifies there are no leaks in the hose

and probe assembly, the pump is operating and the flow fault

circuit is active.

b. With the Eagle in normal operation and with hose and probe

attached, place finger over the end of the probe to block flow.

c. The Eagle should indicate FAIL LOW FLOW LEVEL within a

few seconds.

d. Press RESET SILENCE button to restart pump.
e. If the Eagle does not go into flow fail alarm, inspect hose and

probe for leaks.

f. Have Eagle serviced if you are unable to correct leakage.

Advertising