Confirm CAL C--LIMIT

If the LCD display reads CAL C--LIMIT,

press the Power/Mode button on the gas

monitor to continue to TRANSMIT.

If CAL C--LIMIT is not displayed continue to

the next step.

Calibration Results

Once calibration is complete, the display either shows a “P” for pass or an “F” for fail for each

sensor. The display will alternate between the gas concentration and the letter “P” when the

monitor passes calibration. The “CAL” LED lights in green when calibration is passed. “CAL”

LED lights in red if there is a failure.

Instrument Connected

Once on, the display will show gas channels

and a transmit message. The gas monitor is

ready for a Bump Test or Calibration.

Bump Test

Press the “Bump” button to start bump test.

Note: Ensure calibration gas cylinder is connected

to a demand flow regulator and then to the back

of the cal station. Make sure calibration gas

concentrations match values shown on display.

Bump Gas Reading

During Bump test, the display

shows current gas reading.

Bump Results

The BUMP LED will stop blinking and be on

steadily green if the bump test passed or

steadily red if the bump test failed. Display

will show “P” for pass or “F” for fail for each

sensor. The display will alternate between

the gas concentration and the letter “P”

when the monitor passes bump test.

Return to Transmit Screen

Press the “EDIT/ENTER” button.

See step “E” below for turning off

instrument after completion of

bump test.

Calibrate

Press the “CAL” button to start calibration.

Note: Ensure calibration gas cylinder is connected

to a demand flow regulator and then to the back

of the cal station. Make sure calibration gas

concentrations match values shown on display.

Removing from SDM-2009U

After a successful bump test or calibration,

the GX-2009 will automatically turn off within

15 seconds.

To turn off manually, press

the SDM-2009 “Edit Enter” and “Power”

buttons simultaneously.

Warning: Do not remove the GX-2009

before turning it off, and do not use the

GX-2009’s power button to power it down.

Turn Cal Station On

1.

Press the “POWER” button on the cal

station.

2.

Seat the GX-2009 into the cal station.

Then turn on the GX-2009 by pressing

the “POWER/MODE” button on the gas

monitor.

Calibration Gas Readings

During calibration, the display shows

current gas readings. Typical calibration

will take 3 minutes.

A

B

C

B

D

C

A

E

BUMP

TEST

CALIBRA

TION

DA

TA

TRANSFER

TURN ON

C

A

B

A

D

Data Transfer To USB Flash Drive

With USB flash drive installed and unit powered on,

press and release the COPY button. The COPY LED will

become solid red while the records in the calibration

station’s memory are copied to the flash drive. In

addition, the flash drive’s LED will begin to blink.

When the LED stops blinking, the copy operation is

complete. Remove the flash drive from the USB port.

Transfer data to PC with Calibration Station software,

following the instructions in the Operator’s Manual.