Steriflow SC30 Series User Manual

Page 3

Advertising
background image

S

peCiFiCationS

Sample inlet/outlet connection:

1/2" Tri-Clamp

®

Cooling water connections:

3/4" Tri-Clamp

®

Shell, coil and fittings:

316L Stainless Steel

Coil surface area:

0.97 ft

2

(0.09m

2

)

Coil design pressure:

145 psig @ 365°F (10 barg @

185°C)

Shell design pressure:

145 psig @ 212°F (10 barg @

100°C)

Surface finish

• Wetted surfaces: SF1 20 Ra

µin (

0.5

µm

Ra);

optional: SF5, 20 Ra

µin (

0.5

µm

Ra electropolish)

Note: coil formed from tube having an internal finish
of 10 Ra

µin (

0.25

µm

Ra);

• External body and fittings: satin polish

Available accessories

• Sample valve
• Cooling water hose adaptor kit
• Sample hose
• Carrying case
• Extended Legs

Capacities (approximate)

• Steam – 10 l/h of condensate at 86°F (30°C) from

steam at 43.5 psig (3 barg)

• Water – 30 l/h of water from 185°F to 86°F (85°C to

30°C)

S

ample

C

oolerS

SC30 Sample Portable Cooler

The SC30 Sample Cooler is designed to facilitate
sampling of clean steam and Water For Injection
(WFI), while maintaining product sterility during test-
ing. Constructed in 316L stainless steel, the SC30 is
intended for portable operation at any access point in
those systems.
Sampling is achieved by regulating a flow of clean
steam or WFI through the central coil controlled us-
ing the SV hand valve. A cooling medium, typically
city main water, passes through the shell of the unit
absorbing heat from the test sample, condensing or
cooling the sample prior to discharge from the coil.
All surfaces wetted by the sample medium have a
maximum surface finish of SF1, 20 Ra

µin

(0.5

µin

Ra).

Features offered by the SC30 include:
• 316L stainless steel construction – suitable for

clean steam and WFI systems

• Self-draining design – eliminates possibility of

sample retention

• Fully sterilisable/autoclavable – satisfies valida-

tion criteria

• Availability of hygienic sample valve – allows fine

control of sample flow during testing

• Portable design – single unit can be used to take

samples throughout a system, carrying case
available

19.7

(500

sample inlet

cooling water

inlet

cooling water

outlet

-3-

G

7.9

(200)

6.3 (160)

6.9 (175)

12.0

(305)

11.0

(280)

12.3 (312)

24.0

(610)

Weight: 10# (4,5 kgs)
Note: Foot pads on extended leg

model are 2,4" (60mm) round

with extended legs

Advertising