Guidance and manufacturer’s declaration electromagnetic emissions

The HM LIGHT is intended for use in the electromagnetic environment specified below. The
customer or the user of the HM LIGHT should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment — guidance

Emissions test
CISPR 15

Complies

HM LIGHT

may not be interconnected with other

electrical equipment.

Harmonic
emissions
IEC 61000-3-2

Not applicable

Voltage
fluctuations/ flicker
emissions
IEC 61000-3-3

.
Not applicable

The HM LIGHT is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.

Guidance and manufacturer’s declaration electromagnetic immunity

The HM LIGHT Image Intensifier is intended for use In the electromagnetic environment specified below. The
customer or the user of the HM LIGHT Image Intensifier should assure that it is used in such an environment.

Immunity test

lEC 60601
test level

Compliance
level

Electromagnetic environment —
guidance

Electrostatic discharge
(ESD)
lEC 61000-4.2

±6 kV contact
±8 kV air

±6 kV contact
±8 kV air

Floors should be wood, concrete or
ceramic tile, If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.

Electrical fast
transient/burst
lEC 61000-4-4

±2 kV for power
supply lines

±2 kV for power
supply lines

Surge
lEC 61000-4-5

±1 kV differential
mode

±2 kV common
mode

±1 kV differential
mode

±2 kV common
mode

Mains power quality should be that of a
typical commercial or hospital
environment.

Voltage dips, short
Interruptions and
voltage variations on
power supply Input lines
IEC 61000-4-11

<5 % U

T

(>95 %

dip In U

T

) for 0,5

cycle

40 % U

T

(60 % dip in U

T

)

<5 % U

T

(>95 %

dip In UT) for 0.5
cycle

40 % U

T

(60 % dip in U

T

)

Mains power quality should be that of a
typical commercial or hospital
environment, If the user of the HM LIGHT
Image Intensifier requires continued
operation during power mains
Interruptions, It is recommended that the

Advertising