Replacement of fuse holders, Replacement of main connector – UVP Crosslinkers User Manual

Page 12

Advertising
background image

UV CROSSLINKERS

UVP

12

Note: If more than one tube holder is to be replaced, the wiring schematic diagram may be
of assistance.

Replacement of starters

1.

Locate the starter to be changed. These are positioned in a vertical row on the right hand
side of the internal chassis assembly.


2.

Turn starter(s) 30 degrees anti-clockwise and then pull out of the holder.

3.

Reverse above procedure with recommended starter.


Replacement of starter holder

1.

Locate the starter holder to be changed. It is located on the right hand side of the internal
chassis assembly.


2.

Lever off the starter holder using a flat bladed screwdriver. The holder should now be loose.


3.

Replace unit (part number 36-0005-01). This is achieved by first loosening the internal
reflector which is held in place with eight screws. The screws can be located in the chamber
of the Crosslinker - two on the left, two on the right and four on the top of the chamber.


4.

To replace the starter holder first push the center of the two black spigots, which hold the
starter holder (x2) away from you i.e. toward the Crosslinker chamber. As the reflector has
been loosened, this should be done fairly easily with a small, flat-bladed screwdriver.


5.

The starter can now be relocated.

6.

To secure the starter holder, place the holder in your right hand and with your left hand in the
chamber, press the now loosened reflector adjacent to the holder inwards. This will push the
black spigots inwards thereby securing the starter holder.

7.

Re-fit reflector.

Replacement of fuse holders

These are located at the back of the unit just above the main connector. There are two fuses in
separate holders. Replacement is the same for both.

1.

Cut back the heat shrink on the two wires.


2.

Unsolder both connections and pull from the holder.


3.

Unscrew the nylon nut and push the holder out through the back of the unit.


4.

To replace, reverse the above procedure.

Replacement of main connector

This is located at the back of the unit below the fuse holders.

1. Cut back the heat shrink on the three wires.

2. Unsolder the three connections and pull from connector.

Advertising
This manual is related to the following products: