UVP DE-50 User Manual

Page 3

Advertising
background image

Average erasure time for repeated tests of 50 Intel 2732A EPROMs with recommended erasure dosage of 15
W-sec/cm

2

.



Maintenance

CAUTION: Do not attempt to operate this unit while it is plugged into the AC outlet. Failure to unplug the unit
from the AC outlet before disassembly will result in an electrical shock hazard.

WARNING: Do not attempt to operate this unit in the disassembled condition as that may expose the eyes and
skin to shortwave UV light that is harmful to the unprotected eyes and skin!

This unit is virtually maintenance free. If the lamp fails to start, check to see that:

1. The unit is plugged in.

2. The drawer is pushed all the way in (listen for the click of the limited switch).

3. The timer is on.

4. Check the fuse to see if they are operational.


If a problem still exists, return the unit to the factory for inspection and repair. If the unit has been heavily used
for a long time, it may be desirable to replace the lamp to reduce erase times to their original values.

Tube Replacement Steps


1.

Unplug the unit.

2. Use a Phillip’s screwdriver to remove back cover.

3. Remove drawer by sliding all the way out and then tilting upwards.

4. Using both hands, reach in each side of unit, grasp the tube and gently twist 90

o

(it does not matter which

direction). Tube should easily slip out.

5. Slide new tube into the unit. Using both hands, gently push the replacement tube up and twist into place.

6. Replace unit back cover.

7. Replace drawer (be sure it is fully closed or safely shut-off will prevent unit operation).

8. Plug unit in and observe the viewport for that specific lamp to verify lamp operation.

UVP offers technical support for all of its products. If you have any questions about product use, operation, or
repair, please call or fax UVP Customer Service at:

Corporate Headquarters USA: (800) 452-6788 or (909) 946-3197; Fax: (909) 946-3597

European Operations UK: +44(1)223 420022; Fax: +44(1)223 420561

NOTE: A Returned Goods Authorization (RGA) number must be obtained from UVP before returning any
equipment to UVP.

Replacement Parts List

Description Part

Number


Ballast, 8W, 100V

42-0005-03

Ballast, 8W, 230V

42-0005-04

Ballast, 8W, 115V

42-0005-01

Foam

Pad

72-0049-01

Fuse, 1 Amp

56-0022-02

Fuse

Holder 56-0004-01

Tube,

Shortwave

34-0007-01