UVP ChemiDoc-It Imaging System User Manual

Page 9

Advertising
background image

ChemiDoc-It Imaging System

9

Installation with Manual Lens (BioChemi 510 and OptiChemi 610 Only)

















The photograph shows the parts required for assembly of the OptiChemi 610 camera with manual lens
on the ChemiDoc-It Imaging System. The BioChemi 510 camera will have a different appearance but
will be installed in the same manner.

Note: Not all components may appear as shown in the image.

1. Set up the darkroom and unplug all components from the power outlet.

2. The base mount (3) is preassembled and attached to the ChemiDoc-It darkroom at the UVP

factory. In the case that the base and gasket are shipped separately, secure the base mount
and gasket to the darkroom with the long screws provided (not shown).

3. Attach the camera bracket (4) to the base mount with the brass thumb nuts (10).

4. If not already removed, take the lens cap off the lens. Do not unscrew the lens cap as it will

simply pull straight off.

5. The camera (2) is preassembled with the manual lens (7), step-up ring (6) and diopter (9).

Place the camera/lens assembly over the base mount with the lens tight against the round
cutout in the rubber gasket. Secure the camera to the bracket (4) with the black thumb screw
(5) on the right side of the camera with the power and USB ports facing toward the front of the
unit.

NOTE: The camera shown may appear different than the shipped camera.

6. Connect the camera power supply from the darkroom into the camera. Locate the cable

extending from the top of ChemiDoc-It darkroom.

Note: If upgrading an existing system from a previous camera, do not use the power supply
built into the ChemiDoc-It darkroom; instead, use the external power supply supplied with the
camera kit.

7. Connect the camera’s USB cable to the camera and then into an available port on the

computer.

8. Open VisionWorksLS to operate the camera. Refer to the VisionWorksLS software manual on

the Support CD or use the Help Files embedded in the software for more information.

NOTE: The magnifier (8) shown above is no longer used with the system.

BioChemi 510 &

OptiChemi 610

Components

1. USB Cable

2. Camera

3. Base Mount

and Gasket

4. Bracket

5. Black Thumb

Screw

6. Step-Up Ring

7. Manual Lens

8. Magnifier

9. Diopter

10. Brass Thumb

Nuts

11. Power Cord

12. Camera Power

1

2

3

4

5

6

7

8

9

10

11

12