Troubleshooting

No power to darkroom cabinet.

1. Recheck main power cord connections to the both the EC2 Darkroom and the wall

power (surge protector).

2. Check the fuse located on the main power port. If the darkroom continues to blow

fuses, call UVP Technical Support Department.

Transilluminator will not turn on!

1. Be sure the darkroom cabinet’s door is completely closed. There is a UV exposure

safety cut-off switch that turns the transilluminator of when the darkroom cabinet’s door

is opened.

2. Be sure the darkroom cabinet’s main power switch is lit. If not, refer to “No power to

darkroom cabinet” above.

3. Be sure the transilluminator switch is on. When the switch on the transilluminator is lit

green, the unit is receiving power!

4. Be sure the transilluminator control switch on the darkroom’s front panel is selected to

either “safety switch” or “always on.”

5. Call UVP Technical Support Department with any technical questions.

Replacement Parts

To order replacement parts for the EpiChemi II Darkroom, contact UVP’s offices. Contact

information is listed on the next page.

Part Description

Part Number

Qty.

Tube, 8 watt, fluorescent, cool white

34-0056-01

2

Tube, 8 watt, 365nm UV longwave

34-0006-01

2

Tube, 8 watt, 254nm UV germicidal

34-0007-01

2

Cable and connector wall plug, 115V and 100V

58-0085-01

1

Cable and connector wall plug, 230V

58-0085-03

1