Service Procedures

Care and Cleaning

CAUTION: Warming plate surface can be hot. Let the plate cool prior to cleaning the unit.

To clean the unit:

Unplug the unit from the power source prior to cleaning.

Use mild soap and water with a soft cloth or sponge to clean the exterior of the unit.

Mild isopropyl alcohol or bleach can be used to clean the surface of the unit.

CAUTION: If cleaning agents other than those recommended by UVP are used, a reaction with parts or
materials contained within the warming plate may pose a potential hazard. Contact UVP Technical Support
(contact information listed below) if there are questions regarding the materials used for decontamination or
cleaning.

Return Procedure

A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product or component.

Troubleshooting

No Power to the Darkroom

1. Recheck the main power cord connection to the iBox Explorer

2

darkroom.

2. Check the fuses located at the back of the unit next to

the power port. A small flat-head screwdriver or similar
tool will be required. Push the bottom tab of the
fuseholder up until the fuseholder’s bottom pops out.
Then, push the top tab down until the top pops out.
The entire fuseholder can now be pulled out.

Inspect the thin wire within each glass fuse to see if
there is a break in the wire. If so, replace the fuse(s). If
fuses are blowing repeatedly, contact UVP Technical
Support for additional troubleshooting.

When reinstalling the fuses and fuseholder, make sure that both the bottom and top fuseholder
tabs click firmly into place.