Service procedures, Return procedure, Replacing the battery – UVP Luminescence Calibration Standard User Manual

Page 4

Advertising
background image

LED Calibration Standard

4

Fig. 1

Fig. 2

Fig. 3

Fig. 4

Service Procedures

Return Procedure

A

Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before

returning any product.


Replacing the Battery

The Luminescence Calibration Standard is powered by a standard 9V alkaline battery. To replace the

battery:
1. Locate the Battery Compartment Door on the top of the unit. Remove the Door by removing the screw

holding the door in place (Fig 1).

2. Pull the battery slightly out of the unit (Fig 2). Then, disconnect the Battery Terminal Connector from the

battery (Fig 3).

3. Once the Battery Terminal Connector is fully disconnected, remove the 9V battery from the unit. Then,

slide a new battery in with the positive (smaller) terminal facing toward the front of the unit (Fig 3).

4. Place the Battery Terminal Connector on top of the battery. Press down on the Connector until it snaps

into place on both terminals (Fig 4).

5. Place the Battery Compartment Door back onto the unit, then tighten the screw to hold the door in place

(Fig 1).





Positive Battery

Terminal