UVP J-225 BLAK-RAY User Manual

Page 2

Advertising
background image

HOW TO READ ANGSTROMS - NANOMETERS - ERGs

The U.S. NIST nomenclature requires measurement of wavelength in nanometers (nm)

instead of Angstroms. Ten Angstroms are equal to one nanometer, which also equals

one millimicron. To obtain measurements in ergs/second/cm², multiply µW/cm² by 10.

GENERAL READINGS

If intensity at the surface (such as on a laboratory bench) is to be measured, the sensor

is placed directly on the surface and connected to the meter housing by the extension

cord. Other measurements can be made at varying distances from the source. For

repeatability of readings, the distance should be noted and the sensor placed at the

same distance for each reading. For making measurements at the source, the sensor

face has four pegs for indexing directly on the tube.

The shortwave meter, J-225, is designed for measuring energy from 220nm to 290nm

emitted by germicidal lamps (peak sensitivity is approximately 254nm). The longwave

meter, J-221, measures energy from 300nm to 400nm (peak sensitivity is approximately

365nm) and is basically designed for measuring "black light" lamps.

NOTE: When the sensor cell is positioned directly on the tube, it cannot measure the

ultraviolet from the entire lamp, only the small area it is on. Readings made by this

method are widely used for monitoring lamp condition over a period of time. It should be

noted that these readings are only relative to each other, not a measure of total lamp

energy, as a result of high loss of incident energy due to cosine reflection losses. The

meter cell should be exposed to the UV only long enough to obtain measurement since

extended exposure will degrade cell calibration.

TEMPERATURE/HUMIDITY EFFECTS

It is preferable to use the meters at normal room temperatures (60-90°F). A short

exposure to higher or lower temperatures may be tolerated although accuracy may be

diminished. Upon prolonged exposure to temperatures below the dewpoint, the

instrument may show increased sensitivity (maximum +7%) when immediately used to

make measurements at temperatures above the dewpoint. The cause of the increase in

sensitivity is the condensation of water vapor on the detector filter surface. When the

detector has reached a temperature above the dewpoint and surface condensation has

dissipated, the above specifications will apply.

INFRARED INTERFERENCE

Although the longwave and shortwave sensor cells have their peak sensitivities at

365nm and 254nm, respectively, they do have a low level response to infrared with little

interference or sensitivity to visible light. Since most ultraviolet sources are filtered, the

infrared emission is decreased and the amount of infrared seen by the sensor is

generally less than 5%, well within the accuracy of the meter reading. For instance, the

unfiltered Model B-100A 100 watt mercury vapor BLAK-RAY Lamp has an average

infrared intensity of 10% of the total reading. The value decreases to an average infrared

intensity of 2.6% of the total reading when the

filtered longwave emission is viewed.

For greatest accuracy, the cells should not be used when visible light at the point of

measurement is 40 foot candles or more, or when the equivalent of a 100 watt

incandescent source is within 15 feet. Many high pressure mercury arcs have a high

percentage of infrared emission, as does the sun. For example, the infrared of a G.E.

RS-275 Suntan Lamp typically represents an average of 21.5% of the total longwave UV

reading, as determined by the recommended contrast filter procedure described below.

When such lamps are to be metered, sun ultraviolet studies made, or when ambient

infrared is suspected, the following procedure will provide more accurate ultraviolet

readings.

Accurate ultraviolet measurements with no infrared interference can be achieved by

using the included Clear Snap-On Contrast Filter. Make one reading without the filter,

which will include ultraviolet plus any infrared present. Make another reading by

snapping on the Contrast Filter over the sensor cell. As the contrast filter absorbs all

ultraviolet and the meter sees no visible light, the resultant reading will be infrared only.

When the second (infrared) reading multiplied by 1.06 (an insertion loss correction) is

subtracted from the first reading, the resultant figure will be ultraviolet energy only.

MEASURING HIGH INTENSITY LAMPS

Use the Reduction Screen to measure any high intensity ultraviolet lamp (i.e., a lamp

which drives the needle off the "B" scale). Readings made with the Reduction Screen

are only relative measurements,

not in microwatts/cm², although the Reduction Screen

does multiply the reading by approximately 5 times. To install the Reduction Screen,

simply snap it on to the sensor cell.

TO MEASURE FILTERED SHORTWAVE TUBES AND FILTERS

Use the J-225 shortwave meter. Remove the lamp filter if present and take a reading

directly on the tube. This gives a relative measurement of ultraviolet output without the

condition of the filter affecting the reading. Remember to wear UV protective spectacles

when making this measurement. As shortwave filters deteriorate with use, it is often

desirable to know filter condition. The J-225 meter readily determines transmission. The

best technique is to note the transmission of the new filter held between a shortwave

source and the meter. Subsequent readings on the filter will give the percent degradation

by simple calculation.

The meter sensor cell is subject to the same type of deterioration described above. For

this reason, the meter sensor should be exposed to shortwave ultraviolet light only for

the minimum time required to make the reading. This is particularly important at high

intensities (i.e., those beyond the range of the "A" scale).

MEASURING DETERIORATION OF GERMICIDAL TUBES

The sensor cell should be placed directly on the tube, using the perforated metal

reduction screen clipped on to the sensor. Fit the four pegs on the sensor face over the

tube for proper positioning. This assures the same placement of the sensor each time for

reliable comparative readings. Take a reading approximately four inches from each end

of the tube, and another in the middle of the tube. The average of these three readings

should be used. The reduction screen should be used for all germicidal measurements,

even with low intensity lamps. Remember to wear UV protective eyewear and skin wear

when making this measurement.

CORRECTION FACTORS FOR J-221 METER

A meter measures the density of radiant flux incident on a surface. "Radiant incidence

(irradiance)" is measured in units of power (radiant flux) per unit area, i.e., watts/cm².

Meter measurements yield a number; however, this number will be valid only for a

particular type light source and wavelength range, depending on the design and

calibration of the meter being used.

Advertising
This manual is related to the following products: