UVP UVX Radiometer User Manual

Page 7

Advertising
background image

UVX Radiometer

7

• Place the three position range switch in the 20 mW/cm2 position. Place the READ/HOLD switch

in the READ position.

• Turn the unit on by placing the OFF-ON/TEST switch in the ON/TEST position.

Note: For the first second of operation, the entire display will be turned on demonstrating that all
segments, decimal points and “LOBAT” (the low battery indicator shown on the upper left corner, for
2014 and newer models) on the display are functioning properly. The readout you should see during
that one second period, if everything is functioning correctly, is “LOBAT” (for 2014 and newer models)
and “-18:8.8”. If any one of these segments is not on during this time, refer to the Troubleshooting tips
in this manual.

NOTE: A low battery will be indicated by “LOBAT” text on the display on 2014 and newer models, or
a colon (“:”) for 2013 and older models.

After the one second test mode is completed, the circuit will automatically revert to the normal read
status.

• At this time, the meter should be zeroed to assure the best accuracy. This is done by first

covering the sensor so that no UV light strikes the sensor window. (Generally, turning the sensor
upside down on a bench will suffice.) The range switch is then placed in the 200 µW/cm

2

position

and the zero adjustment trimpot for a display reading of 00.0 µW/cm

2

.

• After setting the zero, ultraviolet readings may be taken by placing the sensor in the UV

environment. The resultant intensities will be shown on the display.