When it is time to measure the UV intensity of the
UV/Air Circulator (recommended weekly or more
often if used frequently), the sensors can be
positioned in front of the sensor access port
located in the center of the module.

Remove the sensor access port by unscrewing
the two thumb screws.

CAUTION: UV/Air Circulator must be unplugged or the power switch turned in the off position to
avoid UV exposure during removal of the access door. Protective eye/facewear and protective
gloves must be worn to avoid exposure to the 254nm UV!

Turn the UV/Air Circulator ON after removing the
access port.

Position the UV sensor over the sensor access
(UVX Radiometer and Sensor shown).

Wait 3 minutes to record the measurement on the
UV meter. It is recommended that the user record
an initial value upon receipt and assembly of the
PCR unit.

After the measurement is taken, reposition the
sensor access door with the thumb screws.

Replacing the UV Germicidal Tube

Turn the power off and unplug the unit from the
outlet.

To change the UV tube in the UV/Air Circulator,
unscrew the four thumb screws from the tube
access door. The tube has two prongs on each end
that fit into a socket. Carefully hold the tube on the
end and insert the tube into the socket. Twist the
tube 90° until the tube locks into place.

Replace the tube access door.

Technical Assistance

UVP offers technical support for all of its products. If you have any questions about the product’s
use, operation or repair, please contact our offices at the locations below. Note: A Returned
Goods Authorization (RGA) number
must be obtained from UVP Customer Service before
returning any product. You may contact UVP’s Customer Service based on your location:

If you are in North America, South America,
East Asia or Australia:

If you are in Europe, Africa, the Middle
East or Western Asia:

Call (800) 452-6788 or (909) 946-3197 regular
business days, 7:00 am and 5:00 pm PST

Call +44(0) 1223-420022 regular business
days, 9:00 a.m. and 5:30 p.m.

E-Mail your message to: [email protected]

E-Mail your message to: [email protected]

Fax Customer Service, and send it to:
(909) 946-3597.

Fax Customer Service, and send it to:
+44(0)1223-420561.

Write to: UVP, LLC 2066 W. 11

th

Street,

Upland, CA 91786 USA

Write to: Ultra-Violet Products Ltd., Unit 1,
Trinity Hall, Nuffield Road, Cambridge
CB4 1TG UK


UV/Air Circulator
tube access

Advertising
This manual is related to the following products: