Care and use manual – Waters ACQUITY UPLC M-Class Columns User Manual

Page 7

Advertising
background image

7

[ CARE AND USE MANUAL ]

Replace the Emitter

Place the pin plug at the back side of the union. This will mark the

center of the union (right).

Thread the replacement emitter (back end first) through the front

of the sprayer and into the union.

Tighten the PEEK nut around the union carefully (do not

over-tighten).

Remove the pin plug and replace it with the transfer line from the

ACQUITY UPLC M-Class Column (again, do not over-tighten
the fitting).

About poorly cut fittings

DO NOT cut or modify the emitter or column tubing.

All tubing and columns supplied by Waters are pre-cut and polished.

Improperly cut tubing can lead to "shedding" and introduction of
fused silica particles downstream of ACQUITY M-Class 75, 100,
or 150 µm I.D.

Columns. This consequently can cause undesired

system backpressure increases.

To reduce the occurrence of clogging

Make sure the union between the column and emitter is clean.

Do not over tighten the fittings on the union (finger tight only).

Make sure the UPLC solvents are clean.

Keep a flow on the tip even when the instrument is not in use

(as low as 200 nL/min).

Try using a TaperTip instead of a Pulled Tip Emitter if

clogging persists.

TaperTip Emitter

A TaperTip (P/N 186003932) is offered as an alternative to a
PicoTip and is the default emitter in the universal sprayer kit.
(July, 2007). The TaperTip has an I.D. of 20 µm but unlike the
PicoTip the I.D. remains unchanged throughout the emitter. This
allows the emitter to be implemented for higher flow rates.
Furthermore, the TaperTip may experience a lower frequency
of particle clogging.

Note: Compared to the PicoTip Emitter, Waters TaperTip Emitter does
not produce as stable a spray at 250-400 nL/min, which is the flow
rate range most customers use with 75 μm or 100 μm I.D. columns.
As indicated above, use of the TaperTip Emitter is the suggested
when a more robust device is preferred.

Column I.D.

Optimal Flow Rate (nL/min)

Emitter Type

75 µm

300

PicoTip

100 µm

400-500

PicoTip/TaperTip

150 µm

1000

TaperTip

300 µm

4000

ESI Source required

ACQUITY UPLC M-Class Column Care & Use