Waters µStyragel Column User Manual

Care and use manual, Μstyragel column

Advertising
background image

[ Care and Use ManUal ]

µStyragel Column

1

Contents

I. IntRoDUCtIon

a. Description

b. Characteristics of µStyragel™ Columns

II. InstallIng the ColUmn

a. Preparing the GPC/HPLC System

b. Connecting the Column to the GPC System

c. Connecting Columns in Series

d. Equilibrating the Column

III. PRePaRIng solVent anD samPles

a. Preparing the Solvent

b. Changing Solvents

c. Preparing the Sample

IV. UsIng the ColUmn

a. Chromatography Guidelines

b. Calibrating the Column

V. CaRe anD maIntenanCe

a. Chromatography Guidelines

b. Calibrating the Column

c. Converting Columns for High Temperature Chromatography

VI. oRDeRIng InFoRmatIon

µstyRagel ColUmn

I. IntRoDUCtIon

This manual covers the use of the Waters µStyragel

TM

family of Gel

Permeation Chromatography (GPC) columns. Please take a few moments to
read this carefully. The recommendations in this manual will help you obtain
good column life and reproducible chromatography.

a. Description

µStyragel columns are packed with high performance, fully porous,
highly cross-linked styrene divinylbenzene copolymer particles. This
column family offers you a choice in moderate to high efficiency, high
resolution, and stable, reproducible operation. Their different resolu-
tion and stability characteristics allow you to choose the optimum
column for your application.

b. Characteristics of µStyragel Columns

Table 1 lists characteristics of the different types of µStyragel
columns.

Table 1: Characteristics of µStyragel Columns

II. InstallIng the ColUmn

a. Preparing the GPC/HPLC System

Before attaching the column in the flow path on a GPC/HPLC system,
you must first prepare the system:

1. Directly connect the system injector to the detector by replacing

the old column(s) with a zero dead volume union.

2. Flush the system to remove any micro particulates and old

solvents. Flush the injector loop if applicable.

3. Remove the union.

Column

Type

Nominal

Particle Size

Shipped in

Avoid using

µStyragel

15 µm

Toluene

Water, alcohols except

HFIP, ketones, or alkanes