[ Care and Use ManUal ]



Part C - How to Apply MassPREP

Enhancer for Highly

Complex Samples

MassPREP

Enhancer can be used to improve the overall selectivity

of the phosphopeptides from a highly complex biological sample (e.g.,
whole cell lysate digested with proteolytic enzymes).

1. Weigh 100 mg Enhancer from the vial and transfer to a separate

vial. Keep the rest of Enhancer in the capped vial for later use, and
store the vial at room temperature. Expiration time of the Enhancer
powder is ~ 3 years; storage is at room temperature.

2. Prepare 100 mg/mL Enhancer solution by adding 1 mL of the loading

solution to the weighed-out enhancer, vortex gently. (Note: The
expiration time for this reconstituted solution is ~ 1 week; storage
is at 4 ˚C)

3. Prepare the peptide sample by solubilizing it with 200 µL of the

Enhancer solution.

4. Precede with Part I to isolate the phosphopeptides.

Part D - Sample Preparation for Mass Spectrometry Analysis

The phosphopeptides enriched using this kit, will be subjected to Mass
Spectrometry analysis. Here are some advices for post-kit sample
preparation.

1. For MALDI-MS analysis, use a reversed-phase cartridge (e.g, ZipTip

)

to remove Diammonium Phosphate or other potential contaminants
from the phosphopeptides.

2. For capillary or nano-scale LC/MS analysis, a reversed-phase trap

column is needed to remove excess EDTA, Diammonium Phosphate
and other contaminants from the sample.

V. ordErIng InforMatIon

Table 2. MassPREP

Phosphopeptide Enrichment Kit Ordering Information

Description

Part Number

Units/ki Amt/Unit

MassPREP

Phosphopeptide Enrichment Kit

186003864

MassPREP

Phosphopeptide Enrichment

µElution Plate

186003820

Qty 1

1 plate

MassPREP

Enolase/Phosphopeptide

Standard

186003286

Qty 1

1 nmole

MassPREP

Enhancer

186003821 Qty 10

500mg


Table 3. Vacuum Manifold Accessories Ordering Information

SPE Vacuum Pump (115, 60 Hz)

725000417

SPE Vacuum Pump (240, 50 Hz)

725000418

µElution Plate Manifold

186001831

Manifold Waste Tray, disposable, 25/pkg

WAT058942

350 µL Sample Collection Plate, polypropylene
96 wells, 50 plates/pkg

WAT058943

VI. StoragE and dISPoSal of uSEd PlatES

Cartridges stored in their original sealed pouch remain stable for long
periods. To store unused cartridges in opened pouches, squeeze the
air out of the pouch, fold over the open end of the pouch twice, seal
with tape and store in a dessicator. Contaminants that may interfere
with the effectiveness of the affinity sorbent are high quantities of
detergents and organic salts, such as urea. Dispose off used cartridges
and plates safely in accordance with applicable government or local
regulations.