Precautions in handling – Yamato Scientific BH500 Thermo-Elite User Manual
Page 9
6
Precautions in handling
8. When you need circulation:
•
Detach flexibility pipe or fitting for
the short-circuit and install the
attached hose joint when you
need to circulation at a distance.
Connect a hose with the hose
joint and secure them with hose
clamps not to leak water. When
using silicon oil, select a hose to
resist silicon oil (e.g. fluorine
rubber).
9. When you use liquid, you should
•
The liquid poured into the examination
tank must use water or the silicon oil.
•
It evaporates for water and alcohol, etc.
while using .Replenish sometimes.
•
Using well water or bad-quality water
will cause accumulation of scale and
copestone on the heater pump, etc. Be
sure to use distilled water or pure water
to prevent performance deterioration or
malfunction.
•
The viscosity must use the one to
50CST at 80
℃ or more about the
silicon oil.
Density of ethanol
Wt %
Vo1 %
Coagulation
point (
℃)
4.8
6.0
-2.0
11.3
14.0
-5.0
13.8
17.0
-6.1
16.4
20.2
-7.5
17.5
21.5
-8.7
18.8
23.1
-9.4
20.3
24.8
-10.6
22.1
27.0
-12.2
24.2
29.5
-14.0
26.7
32.4
-16.0
29.9
36.1
-18.9
33.8
40.5
-23.6
39.0
46.3
-28.7
46.3
53.8
-33.9
hose joint
Fitting
Flexibility pipe for the short-circuit
hose