4.

Installation/Assembly

Assembling parts

1. Set the Body at a stable place.

Be sure to set the body at a stable place.

2. Attach the motor to the body.

1. Insert the motor bearing bar into the motor support clamp of the body, put D cut surface (flat

surface) of the bar perpendicularly to either 2 upper or side screws, and fasten 4 screws tightly by

using the attached hexagonal wrench (for M5).

Then, slant the motor to the right (about 45°), and finger tightly the motor screw.

If you do not fasten the screws tightly, vibration may occur preventing accurate

measurements or the motor may fall causing the glass apparatus to break.

2. Remove the cooling condenser nut (the bigger nut with the coil ring) and coil ring when you

attach the motor.

Motor Screw

Motor Support Clamp