In the Event of Failure…

Safety Device and Error Code


This unit has an automatic diagnosis function built in the controller and safety devices independent of the
controller. The table below shows the cause and the solution method when the safety device operates.

Error Code:

If an error in use or equipment a failure occurs, the temperature display screen on the operating panel
displays the corresponding error code and the alarm buzzer sounds. The buzzer stops by pressing any key.
In case an error occurs, check the error code and turn off the earth leakage breaker.

Safety Device

Notify

Cause/Solution

Sensor trouble detection

Er.01” appears

• Failure in temperature input circuit.
• Temperature sensor is broken or disconnected.
• The measured temperature is out of the display range.
™

Make a call for service.

SSR short-circuit
detection

Er.02” appears

• SSR is in short-circuit
™

Make a call for service.

Heater disconnecting
detection

Er.03” appears

• Heater is disconnected.
™

Make a call for service.

Cover locking error

Er.04” appears

• The cover is unlocked during operation.
™

Make a call for service.

Cover unlocking error

Er.05” appears

• The cover is not unlocked at releasing.
™

Make a call for service.

Incorrect bottle position

Er.06” appears

• Bottle is attached incorrectly.
™

Open the door and attach the drain bottle correctly.

Overheat error

Er.07” appears

• The chamber temperature rises to 140℃ or above.
• The temperature of "preset temperature + (plus) 3℃

or above" is continued for one minute during
sterilization process.

• The temperature of "preset temperature + (plus) 10℃

or above" is continued for ten minutes during
heat-retention process.

™

Make a call for service.

Sample sensor (optional)
disconnection

Er.08” appears

• Disconnection or abnormality of sample sensor.
• The setting is set to "on" when the sensor is not

attached.

™

Check the setting referring to "Maintenance Mode" in

Page 31.

™

Make a call for service.

Exhaust valve error

Er.09” appears

• Failure in exhaust valve.
™

Make a call for service.

A/D conversion error

Er.14” appears

• Failure in electrical parts.
™

Make a call for service.

Memory error

Er.15” appears

• Failure in preset value of memory.
™

Make a call for service.