1. Specifications

Time-up Output

Output at the automatic stop and at the end of program

C contact (Relay contact)

Contact capacity

:AC250V 3A

      :DC30V 3A

Joint

:M4 Screw terminal (Common)

Alarm Output

Output at the occurrence of error

C contact (Relay contact)

Contact capacity

:AC250V 3A

      :DC30V 3A

Joint

:M4 Screw terminal (Common)

Temperature Output

(Analog)

Voltage output(DC) according to the measured temperature

See table 1 in the next page about output

(0

℃ at 0V)

Resolution

:0.1

Load impedance

:100K

Ω or more

Joint

:M4 Screw terminal (Common)

2. Connection

1. Connect firmly the terminal to use (Outputs which was not specified at the order are

unusable)

2. To use the time-up output or the alarm output;

Connect to COM (C) and NO in case the contact is “Closed” at output.

Connect to COM (C) and NC in case the contact is “Opened” at output.

3. when you use the temperature output, employ a shielded line as a connecting cable for

noise prevention, and cut it as short as possible.

Lengthwise Arrangement

Crosswise Arrangement