Measurements

59

Time Domain (Zero Span)

In Time Domain, the amplitude of a single frequency is measured,

rather than sweeping a range of frequencies. The SignalHawk mea-

sures and displays the amplitude of the frequency for a specified

period (sweep time) and refreshes during the next sweep. The Time

Domain trace resembles the horizontal line display on an oscilloscope.



Note:

The center frequency will being measured.

1. Activate the measurement by doing one of the following:

z

Select Time Domain in the Start Menu.

z

Select Time Domain in the Measurement Menu.

z

Set the Frequency Span to 0 Hz. See “Freq & Span Menu” on

page 31.

2. Press the Sweep Time soft key and enter how long a single sweep

should be using one of these methods:



Note:

Time range can be set from 1 ms to 100 s.

z

Up/Down Arrows: Step increments of one second.

z

Right/Left Arrows: Step increments of one millisecond.

z

Mouse Scroll Wheel: Step increments of one second per click.

z

Number Keys: Enter the time in seconds.



Note:

There are a limited number of pixels, so longer

sweep times mean less detail in the display.



Note:

This enables the Video Trigger in the BW & Sweep

menu. To use the Video Trigger, see “Resolution & Video BW”

on page 36.