Precision power sensors, 4027a series, Specifications – Bird Technologies 4027A Series-Datasheet User Manual

Page 2

Advertising
background image

Precision Power Sensors

4027A Series

PowerSensor_4027A-09202012

30303 Aurora Rd.

Solon, OH 44139

866.695.4569

www.bird-technologies.com

SPECIFICATIONS

Accuracy ±1% at calibration frequencies and power levels

±2 % at other frequency and power levels
Add 2% to uncertainty outside 25 ± 10 °C

Calibration

Power Level

1000W units: 700 watts
10kW units: 1700 watts

Uniformity 2 % maximum unit to unit, at calibration

frequency and power levels

Speed 2 readings per second

Maximum Power 10 kW units - 12 kW max.

1 kW units - 1.2 kW max.

Harmonic Content -50 dBc max

VSWR Range 1.0-2.0

Directivity 28 dB

Insertion Loss <0.05 dB

Connectors *Customer Specified

Power Requirements

External DC

12 VDC, supplied from Bird 4421 Power Meter

Dimensions 5.2” L x 2.5” W x 3.25” H

Weight 1 lbs. 13 oz. (0.8 kg)

Operating Temperature 15°C to 35°C (59°F to 95°F)

Storage Temperature -40°C to 80°C (-40°F to 176°F)

Humidity 95% maximum (non-condensing).

Altitude Up to 10,000 feet (3,048 m)

General EMC Designed to carry CE mark (with immunity

exception noted below)

Emissions EN-55011, 1991, Class B

Immunity EN-50082-1, 1995

Safety EN-61010, 1993 in accordance with Council

Directives 73/23/EEC and 93/68/EEC

Calibration Cycle

6 month. Performance before and performance

after data to be supplied for units

If you need assistance in selecting products from our standard 4020-series sensor line,
please contact a sales engineer at Bird Technologies Group.

*For connector options, please refer to our catalog or contact sales at 866.695.4569 /
sales@bird-technologies.

Models

Power Range

Frequency

4027A12M

300 mW to 1 kW

10-15 MHz

4027A250K

3 W to 10 kW

250-400 kHz

4027A400K

3 W to 10 kW

400-550 kHz

4027A800K

3 W to 10 kW

800-950 kHz

4027A2M

3 W to 10 kW

1.5-2.5 MHz

4027A4M

3 W to 10 kW

3-5 MHz

4027A10M

3 W to 10 kW

10-15 MHz

4027A25M

3 W to 9 kW

25-30 MHz

4027A35M

3 W to 7.5 kW

35-45 MHz

4027A60M

3 W to 6 kW

45-65 MHz

4027A100M

3 W to 4 kW

95-105 MHz

4027A150M

3.75W to 3.75 kW

150-170 MHz

Also available - Standard 4020 series 4021(300mW-1kW, 1.8-32 MHz),
4022 (300mW-1kW, 25-1000 MHz), 4024 (3W-10kW, 1.5-32 MHz),
and 4025 (3W-10kW, 100-2500 kHz).
±3% (1s) of reading accuracy and 28 dB minimum directivity.

Note 1: See also 4020 Series of broadband, 5% accuracy sensors.
Note 2: For applications with harmonic greater than -50 dBc,
contact the factory for versions of 4027A Sensors with filtering
included.

PowerSensor_4027A_2012 branding 7/30/2013 2:58 PM Page 2