background image

4421 Instructions

10

Error Codes

The Bird 4421 displays error codes when the RF power is either below the
selected range (underrange) or above the selected range (overrange). Figure 6
displays the error codes and Figure 7 lists the function limits.

Figure 6 Error Codes

Figure 7 Function Limits

Audible Warning

If the RF power level exceeds 120% of the sensor’s power range, the meter will
sound a warning buzzer.

Symbol

Explanation

Value greater than overrange limit of function

Value less than underrange limit of function

Function

Limit

Error

FWD, RFL

Power > 199.9% of full scale or 120% of top
range

Overrange

FWD dBm, RFL dBm

Power > 120% of full scale
Power < 3% of low range

Overrange
Underrange

SWR

FWD < 20% of low range
FWD – RFL = 0

Underrange
Overrange

Return Loss

FWD < 20% of low range
RFL < 20% of low range
Return Loss > 40 dB

Underrange
Underrange