Chapter 3 operation, Zeroing sensor power measurement accuracy, Accuracy – Bird Technologies 5015 Series-Manual User Manual

Page 17

Advertising
background image

5

Chapter 3

Operation

Zeroing Sensor

Over time, the sensor’s “zero value” (reading with no
applied RF power) can drift, making all readings inac-
curate by this value. For example, if the zero value is –
0.02 W, measuring a 50 W signal will give a reading of
49.98 W, a 0.04% error. Measuring a 1 W signal will
give a reading of 0.98 W, a 2% error. If the drift would
be a significant error, rezero the sensor:
1. Ensure the sensor has reached a stable operating

temperature.

2. Ensure no RF power is applied to the sensor.
3. Press “Zero” to begin Calibration.

Note:

Calibration will take about 40 sec-

onds. Do not interrupt the calibration! A bar on
the screen will display calibration progress.

Power Measurement Accuracy

Accuracy

The Bird power sensors are highly accurate. Accuracy is
specified for each sensor type is typically given as a per-
cent of reading or of full-scale.

Advertising
This manual is related to the following products: