Replacing the Instrumentation Module

1.

Remove the back as above in meter replacement.

Note:

The instrumentation module contains the line section and cir-

cuit board chassis as an integral part.

2.

Loosen the dress nut on the CW/PEAK switch.

Note:

Care should be taken not to scratch the label.

3.

Remove the wire assemblies from the batteries, meter, and DC jack.

4.

Remove the battery wires (item 4) first at the printed circuit board by pull-
ing up on the slip-on connector.

5.

Remove the DC jack wires (item 5).

6.

Remove the meter wires (item 3) also at the printed circuit board.

Note:

Care should be taken not to break the wires. \

Note:

Note the wire location for ease of reassembly.

7.

Remove the two oval head Phillips screws located on the front face of the
unit on either side of the line section element port.

8.

Pull the instrumentation module straight out from the back of the unit.

9.

Replace the connector.

10. Zero adjust the meter.

CAUTION

If other than Female N type connectors are used, limit power and frequency to

the capabilities of the RF coaxial cable or connectors used. Damage to

connectors or errors in reading could result.