Connecting Sensor

Although unlikely, it is possible to corrupt the firmware in the power

sensor (DPS) by connecting it to a power meter, such as the Bird

Model 5000 (DPM) or a Bird Site Analyzer (SA), while the power

meter is turned on. To prevent this, make sure that the power meter is

turned off before connecting it to the power sensor. When you are done

making measurements, turn the power meter off before disconnecting

the power sensor.

Element Orientation

The element in the Forward socket should be inserted with its arrow

pointing in the direction of forward power, as shown below. The

element in the Reflected socket should be inserted with its arrow

pointing in the direction of reverse power. The forward element must

have a power rating 10x the reverse element power rating for

optimum results.

CAUTION

Do not connect or disconnect sensor interface cable to the DPS

while the DPM or SA is on. Always turn off the power meter before

connecting or disconnecting a sensor.

WARNING

Leaking RF energy is a potential health hazard.

Never attempt to connect or disconnect equipment from the

transmission line while RF power is being applied. Severe burns,

electrical shock, or death can occur.