User's manual for opto-rs cable connections – Bowers Group Sylvac Opto Instruments User Manual

Page 13

Advertising
background image

12

User's manual for OPTO-RS cable connections

General

The OPTO-RS cable enables a direct connection with most of the Sylvac measuring instruments to a personal
computer, a dedicated printer or to a Sylvac display unit.

It is not only a cable, but an interface which converts the data output of the instrument to a compatible RS232
signal. The periphery instrument connection must be able to supply power to the OPTO-RS plug.

Definition

RS232 communication parameters

4800 baudrate, even parity, 7 data bits, 2 stop bits

Data format

Data

[ Sign ¦ E1-En ¦ "." ¦ F1-Fn ¦ CR ]

Sign : « + », « - », or « space »
E1-En: integer
F1-Fn: fractional
n: depends on used unit and resolution

Errors

[ "ERR" ¦ Number ¦ CR ]

0: sensor error (e.g. speed, scale distance)

1: incorrect command

2: parity error (duplex instruments only)

3: exceeded measurement range

Identification

[ "SY" ¦ Instr. ¦ "." ¦ OPT1 ¦ {"." ¦ OPT2} ¦ CR ]

SY: Sylvac
Instr: 203, 235, 233, etc
OPT1: version option
OPT2: additional version options (according to instrument used)
Note: The id. transmission is done only when switching ON the instrument

Connection description

Two different types of OPTO-RS plug connections are available: Simplex and Duplex

Simplex cable

First generation of OPTO-RS cable connection, designed for instruments which were not able to receive RS232
commands. Data requests are made by LED status change (e.g. by turning off the DTR signal line for a
minimum of 110ms).
The OPTO-RS simplex cable can also be used with duplex instruments, however remote commands will be
ignored.